Inquiry

MULP40

ORANGE SERIES
MULP40

The EOS Power Ultra Low Profile Orange Series sets new height profile benchmarks in the power supply industry with a maximum height of just 0.75 inches, that’s an incredible 19.05 mm! The MULP Orange Series packs up to 30 watts per cubic inch into their small profiles which enables customers miniaturize their applications without compromising on available power levels.

FEATURES
  • 3 x 2 x 0.75 Inches Form factor
  • 40 Watts Convection
  • Approval to EN60601 3rd Edition
  • Efficiencies 85 % Typical
  • -40 to 70 degree operating temperature*
  • Dual fusing
  • 2m Hours, Telcordia -SR332-issue 3 MTBF < 0.3 W
  • No Load Power < 0.3 W
  • Medical (BF) Safety Approvals
  • Meets standard IEC60601-1-2 : 2014 (4th Edition)
AVAILABLE WITH COVER KIT
MULP40_Coverkit

ULP40-CK Series


Input Specifications
Input Voltage 85-264 VAC/390 VDC, Universal                                                                       
Input Frequency 47-63 Hz
Input Current 0.8 A max @ 115 VAC AND 0.4 A max @ 230 VAC
No Load Power < 0.3 W typical
Inrush Current 25 A @ 115 VAC,  45 A @ 230 VAC,  75 A @ 264 VAC
Output Specifications
Output Voltage From 5 V to 58 V (Refer model number matrix table)
Output Power Up to 40 W Convection Cooling
Hold-up Time 40 W : 60 ms @ 230 VAC
Line Regulation +/-0.5 %
Load Regulation +/-1 %
Minimum Load 0.0 A
Transient Response max excursion 4 %; 25 % step load change @ 0.1 A/us slew rate, 50 % duty cycle, 50/60Hz, recovery time < 5 ms.
Ripple1 1.0 % max for all outputs, 1.5 % max for 5 V
Rise Time 50ms typical
Set Point Tolerance 2 % (3 % for 5V model)
Over Current Protection >110 %
Over Voltage Protection 110 to 140 %
Short Circuit Protection Hiccup mode
General Specifications
Efficiency 85 % Typical                                                                                                                                     
Isolation Voltage Input to Output – 4000 VAC medical applications.
Input to GND - 1500 VAC (Not Applicable For Class II Option)
Output to GND- 1500VAC for type BF , 500 VAC for type B (Not Applicable For Class II Option)
MTBF 2m Hours, Telcordia -SR332-issue 3
Switching Frequency 65 KHz Typical
Leakage Current 300 uA typical, (N.A. for Class II Option); touch current <100 uA
Dimensions 3 x 2 x 0.75 inches
(76.2 x 50.8x 19.05 mm)
Weight 100 gm approx
Environmental Specifications
Operating Temperature -40 to 70°C, (-40 to 0°C startup is guaranteed with spec deviation)2                                                                 
Storage Temperature -40 to +85°C
Relative Humidity 5 % to 95 %, noncondensing
Altitude Operating: 16,000 ft.; Non-operating: 40,000 ft
Model Number Power Voltage Max. Load
MULP40-1X05 40 W 5 V 5.00 A
MULP40-1X12 40 W 12 V 3.33 A
MULP40-1X15 40 W 15 V 2.67 A
MULP40-1X24 40 W 24 V 1.67 A
MULP40-1X30 40 W 30 V 1.33 A
MULP40-1X48 40 W 48 V 0.83 A
MULP40-1X58 40 W 58 V 0.69 A
For Screw Terminal version replace "X" above with "0", example MULP40-1005
For PCB mountable version replace "X" above with "2", example MULP40-1205
For Header version replace "X" above with "3", example MULP40-1305
Cover Kit available ULP40-CK
General Notes:
  • Specifications are for nominal input voltage, 25°C unless otherwise stated.
  • When used in Cover Kit, de-rate output power to 70 % under all operating conditions.
  • Class II version available, Add "-II" suffix at the end of the Model Number.
Notes:
  1. Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Electrolytic capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges.
  2. Output ripple can be more than 10 % of the output voltage.
Derating Curve
ULP40
De-rate linearly from 100 % at 50°C to 50 % at 70°C
EMC
Parameter Conditions/Description Criteria
Conducted Emissions EN 55011-B,CISPR22-B, FCC PART15-B Pass
Radiated Emissions EN 55011 A Pass
Level B with external core (King core K5B RC 25x12x15-M in input cable)
Input Current Harmonics EN 61000-3-2 Class A
Voltage Fluctuation and Flicker EN 61000-3-3 Pass
ESD Immunity EN 61000-4-2 Level 4, Criterion A
Radiated Field Immunity EN 61000-4-3 Level 3, Criterion A
Electrical Fast Transient Immunity EN 61000-4-4 Level 3, Criterion A
Surge Immunity EN 61000-4-5 Level 3, Criterion A
Conducted Immunity EN 61000-4-6 Level 3, Criterion A
Magnetic Field Immunity EN 61000-4-8 Level 4, Criterion A
Voltage dips, interruptions EN 61000-4-11 Criterion A & B
Safety
CE Mark Complies with LVD Directive
Safety Standard(s) EN60601-1, IEC 60601-1 (ed.3), ANSI / AAMI ES 60601 - 1, CSA C22.2 No. 60601-1
Approval Agency Nemko, UL, C-UL
Safety File Number(s) Class-I : UL: Certificate No: 2017-1-30-E173812, Nemko: Certificate No: P17221669, CB Test Certificate No : NO95698
Class-II : UL: Certificate No: 2017-01-31-E173812,Nemko: Certificate No: P17221705, CB Test Certificate No: NO95847
Downloads
CB Class - I Certificate
CB Class - II Certificate
NEMKO Class - I Certificate
NEMKO Class - II Certificate
UL Class - I Certificate
UL Class - II Certificate
Mechanical Specifications
AC Input Connector (J1) Option 1 & 3 Tyco: 640445-3
Mating: 647402-3; Pins: 3-647409-1
Pin 1
Pin 2
Pin 3
AC LINE
NOT FITTED
AC NEUTRAL
DC Output Connector (J2) Option 1 & 3 Tyco: 640445-4
Mating: 647402-4; Pins: 3-647409-1
Pin 1,2
Pin 3,4
V1 +VE
V1 -VE
Tyco Header
MULP40
PCB Mountable
ULP40
Screw Terminal
ULP40
    General Notes: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following
  1. Stand off, used to mount PCB has OD of 5.4 mm max.
  2. Screws, used to fix PCB on stand off, have head dia of 6.0 mm max.
  3. Washer, if used, to have dia of 6.5 mm max.
Safety Certificate
CB Class - I Certificate
CB Class - II Certificate
NEMKO Class - I Certificate
NEMKO Class - II Certificate
UL Class - I Certificate
UL Class - II Certificate

Related Products